DIGITAL MULTIMETER DT4282
Rated 0 out of 5
S$ 783.70 *Pricing inclusive 9% GST Add to cart

DIGITAL MULTIMETER DT4282

S$ 783.70 *Pricing inclusive 9% GST

World’s Premier Digital Multimeter. Superior Accuracy and High Response, Topped with Safety Terminal Shuttles

Key features
• 60000 count, 5-digit display, high-resolution measurements
• ±0.025% DC V basic accuracy, wide 20 Hz to 100 kHz AC V
frequency characteristics
• Low-pass filter cuts high harmonics (when measuring inverter fundamental waveforms)
• Includes multiple measurement functions such as DC+ACV, temperature, capacitance, and frequency
• Terminal shutter mechanism (prevents erroneous test lead insertion)
• Measure up to 10A with direct input (DT4282 only)
• Dual display lets you check voltage and frequency simultaneously
• Magnetic strap (Optional)
• Rear kickstand
• Store probes at the back of the tester
• Identify excessively high input with a red screen backlight
• Robust design capable of withstanding a drop from a height of 1 m
• USB communications function supports PC measurements (optional)
• Broad -15 (5°F) to 55°C (131°F) operating temperature range

Available on backorder

Compare
  • Price shown above is the price before GST
  • Shipping cost is not included

The Hioki DT4200 Digital Multimeter Series consist of a complete lineup of 9 models ranging from professional to industrial to pocket types. The DT4282 is a high-precision 60,000 count digital multimeter with direct 10A input and all of the functions and features required for advanced testing in laboratories and R&D.

Additional information

Weight 0.865 kg
Dimension

129W×205H×67D mm

Applications

General Measurement

Nos. Channel

1

Safety

CAT III 1000V, CAT IV 600V

Resistance

600MΩ

AC voltage range

1000V

DC voltage range

1000V

Specifications

Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Regarding DMM Accuracy Due to the many ranges and functions available in a DMM, only the basic accuracy is indicated for reference. Please refer to the individual catalogs for detailed accuracy information.

DC Voltage range 60.000 mV to 1000.0 V, 6 ranges, Basic accuracy: ±0.025 % rdg. ±2 dgt.
AC Voltage* range 60.000 mV to 1000.0 V, 6 ranges, Frequency characteristics: 20 Hz – 100 kHz
Basic accuracy 45 – 65 Hz : ±0.2 % rdg. ±25 dgt. (True RMS, crest factor 3)
DC + AC Voltage* range 6.0000 V to 1000.0 V, 4 ranges, Frequency characteristics: 20 Hz – 100 kHz
Basic accuracy 45 – 65 Hz : ±0.3 % rdg. ±30 dgt. (True RMS, crest factor 3)
Resistance range 60.000 Ω to 600.0 MΩ, 8 ranges, (Conductance: 600.00 nS, DT4282 only)
Basic accuracy: ±0.03 % rdg. ±2 dgt.
DC Current range 600.00 μA to 10.000 A, 6 ranges
Basic accuracy: ±0.05 % rdg. ±5 dgt.
AC Current* range 600.00 μA to 10.000 A, 6 ranges,
Basic accuracy 45 – 65 Hz : ±0.6 % rdg. ±5 dgt. (True RMS, crest factor 3)
Frequency characteristics: 20 Hz – 20 kHz (at 600 μA to 600 mA range)
AC Current* range (use with Clamp on probes) N/A
Peak DC V measurement: Signal width 4 msec or more (single), 1 msec or more (repeated)
AC V, DC/AC A measurement:
Signal width 1 msec or more (single), 250 μsec or more (repeated)
Capacitance range 1.000 nF to 100.0 mF, 9 ranges, Basic accuracy: ±1.0 % rdg. ±5 dgt.
Continuity Check Continuity threshold: 20/50/100/500 Ω, Response time: 10 ms or more
Diode test Open terminal voltage: 4.5 V or less, Testing current 1.2 mA or less, Threshold of forward voltage: 0.15 V to 3 V, seven stages
Frequency range AC V, DC+AC V, AC A measurement, at pulse width 1 μs or more (50 % duty ratio)
99.999 Hz (0.5 Hz or more) to 500.00 kHz, 5 ranges, ±0.005 % rdg. ±3 dgt.
dB conversion Standard impedance setting (dBm), 4 Ω to 1200 Ω, 20 stages
Display dB conversion value of AC voltage (dBV)
Temperature (thermocouples) K: -40.0 °C to 800.0 °C (-40.0 °F to 1472.0 °F)
Add accuracy of the Thermocouple probe to main unit accuracy: ±0.5 % rdg. ±3 °C
Other functions Filter function (Remove harmonic noise, use only at 600 VAC, 1000 VAC ranges), Display value hold, Auto hold, Max/Min value display, Sampling select, Relative display, Measurement memory (400 data), Auto-power save, USB communication (option), 4-20 mA % conversion
Display Main and Sub displays: 5-digits LCD, max. 60000 digits
Display refresh rates 5 times/s (Capacitance measurement: 0.05 to 2 times/s, depending on measured value, Temperature: 1 time/s )
Power supply LR6 (AA) alkaline batteries ×4, Continuous use: 100 hours
Dimensions and mass 93 mm (3.66 in) W × 197 mm (7.76 in) H× 53 mm (2.09 in) D, 650 g (22.9 oz) (with test leads holder and batteries)
Accessories Test Lead L9207-10 ×1, Instruction Manual ×1, LR6 Alkaline Battery ×4